Performance Evaluation of Coupling Variants for an Active Middle Ear Implant Actuator: Output, Conductive Losses, and
Ute A Gamm1,2, Nils K Prenzler2, Max E Timm2
1Cochlear Deutschland GmbH & Co. KG.
Introduction:
This study aims to investigate the performance of an active middle ear implant actuator for various coupling configurations. Actuator output and conductive losses were measured, and the stability of coupling was evaluated by challenging the link between actuator and ossicles through pressure events in magnitudes that occur in daily life.
Methods:
Actuator coupling efficiency and the occurrence of conductive losses were measured in 10 temporal bones through laser Doppler vibrometry on the stapes footplate for various coupling types (incus short process with and without laser hole, incus long process, stapes head). To test the stability of coupling, actuator output was measured before and after daily-life pressure events that were simulated; Valsalva maneuvers (500 cycles of -40 to +60 hPa) and jumping into a swimming pool and diving 3 m deep (a step change of 300 hPa).
Results:
Actuator output was similarly high for all types of coupling to the incus (short process and long process) and most efficient for coupling to the stapes head. Conductive losses occurred in two temporal bones (TBs) for short process coupling but for seven TBs for coupling to the incus long process. All coupling types were stable and did not lose efficiency after pressure events in the low-frequency range (<1 kHz). Losses in output of 13 to 24 dB were observed in one TB at frequencies from 3 to 6 kHz.
Conclusion:
Actuator output was similarly high for all types of coupling to the incus but coupling to the incus long process led to a higher occurrence of conductive losses. All three coupling configurations connected the actuator securely to the ossicular chain, under variations of barometric pressure that can be expected in daily life.
More Related Videos
03:55Author Spotlight: Enhancing Grasping Abilities for Hemiplegic Patients with Flexible Robotic Limbs
Published on: October 27, 2023
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...