Development and validation of an LC-MS/MS generic assay platform for small molecule drug bioanalysis
Robert A Parise1, Joseph M Covey2, Melinda G Hollingshead3
1Cancer Therapeutics Program, UPMC Hillman Cancer Center, Pittsburgh, PA, United States.
Aim:
We developed a generic high-performance liquid chromatography mass spectrometry approach for quantitation of small molecule compounds without availability of isotopically labelled standard.
Methods:
The assay utilized 50 microL of plasma and offers 8 potential internal standards (IS): acetaminophen, veliparib, busulfan, neratinib, erlotinib, abiraterone, bicalutamide, and paclitaxel. Preparation consisted of acetonitrile protein precipitation and aqueous dilution in a 96 well-plate format. Chromatographic separation was achieved with a Kinetex C18 reverse phase (2.6 microm, 2 mm x 50 mm) column and a gradient of 0.1 % formic acid in acetonitrile and water over an 8 min run time. Mass spectrometric detection was performed on an AB SCIEX4000QTRAP with electrospray, positive-mode ionization. Performance of the generic approach was evaluated with seven drugs (LMP744, olaparib, cabozantinib, triapine, ixabepilone, berzosertib, eribulin) for which validated assays were available.
Results:
The 8 IS covered a range of polarity, size, and ionization; eluted over the range of chromatographic retention times; were quantitatively extracted; and suffered limited matrix effects. The generic approach proved to be linear for test drugs evaluated over at least 3 orders of magnitude starting at 1-10 ng/mL, with extension of assay ranges with analyte isotopologue MRM channels. At a bias of less than 16 % and precision within 15 %, the assay performance was acceptable.
Conclusion:
The generic approach has become a useful tool to further define the pharmacology of drugs studied in our laboratory and may be utilized as described, or as starting point to develop drug-specific assays with more extensive performance characterization.