Calculating Heart Rate Variability from ECG Data from Youth with Cerebral Palsy During Active Video Game Sessions
Published on: June 5, 2019
The validity and reliability of an open source biosensing board to quantify heart rate variability
Joel S Burma1,2,3,4,5,6,7, Andrew P Lapointe6,7,8,9, Ateyeh Soroush7,8,9
1Cerebrovascular Concussion Laboratory, University of Calgary, Alberta, Canada.
This study validates the OpenBCI cyton board for heart rate variability (HRV) assessment, showing it is a reliable and cost-effective alternative to expensive electrocardiogram (ECG) systems for quantifying autonomic function.
Area of Science:
- Biomedical Engineering
- Physiological Monitoring
- Wearable Technology
Background:
- Heart rate variability (HRV) quantifies autonomic function but typically requires costly electrocardiogram (ECG) equipment.
- Existing HRV measurement systems are often expensive and inaccessible for widespread research or clinical use.
Purpose of the Study:
- To determine the validity and reliability of an open-source OpenBCI cyton biosensing board for accurate HRV quantification.
- To assess the feasibility of using a low-cost, open-source system for autonomic function assessment.
Main Methods:
- A 3-lead ECG configuration using the OpenBCI cyton board was utilized to record heart rate waveform data.
- Concurrent validity was assessed by comparing the cyton board's HRV metrics against an industry-standard ADInstruments ECG (n=15).
- Between-day reliability was evaluated over a 7-day period (n=10), with HRV metrics analyzed in time- and frequency-domains.
Main Results:
- The OpenBCI cyton board demonstrated excellent agreement and high association (r² > 0.98) with the industry-standard ECG.
- No significant differences in time- or frequency-domain HRV metrics were found between the two devices.
- The cyton board exhibited excellent between-day reliability (Cronbach's Alpha ≥ 0.916, ICC ≥ 0.930) and low measurement error.
Conclusions:
- The open-source OpenBCI cyton board is a valid and reliable tool for multimodal HRV assessments.
- This system offers a significantly more affordable alternative (~4% of standard ECG setup cost) for quantifying autonomic function.
- The cyton board's reliability supports its application in longitudinal studies and broader research settings.
More Related Videos
Related Concept Videos
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Special considerations while measuring pulse
Regulation of Heart Rates
The SNS increases heart rate through the release of norepinephrine and epinephrine, which act on beta-1 adrenergic receptors in the heart. This action increases the rate of depolarization in the sinoatrial (SA) node, the heart's...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.