Validation of the BPro radial pulse waveform acquisition device in pregnancy and gestational hypertensive disorders
Adam Jakes1, Julie Wade1, Zoe Vowles1
1Maternity Services, Guy's and St. Thomas' Hospital NHS Trust, St. Thomas' Hospital.
Objective:
To validate the BPro blood pressure (BP) wrist device for use in pregnancy and preeclampsia according to the Universal Standard protocol.
Participants And Methods:
BP was measured sequentially in 45 pregnant women (including 15 with preeclampsia, 15 with gestational hypertension and 15 who remained normotensive) alternating between a mercury sphygmomanometer and BPro device.
Results:
The BPro is accurate in pregnancy with a mean device-observer difference of -1.7 ± 6.1 and 0.1 ± 4.6 mmHg for SBP and DBP, respectively. In women with preeclampsia, BPro also met the validation criteria for the Universal Standard protocol with a mean device-observer difference of -2.7 ± 7.1 and 0.3 ± 4.7 mmHg for SBP and DBP, respectively. However, the number of absolute BP differences within 5 mmHg was considerably fewer in those with preeclampsia when compared to the other two subgroups.
Conclusion:
The BPro device can be recommended for BP measurement in pregnancy but should be used with caution in those with confirmed preeclampsia.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.