Three-dimensional mapping for His-bundle pacing in pediatric patients, a case series
Anastasia Bury1, Daniel Cortez2
1Central Michigan University College of Medicine, Mount Pleasant, USA.
Journal of Electrocardiology
|June 15, 2021
Summary
His bundle pacing using 3D mapping offers a promising, low-fluoroscopy alternative for pediatric pacemaker patients. This technique demonstrated successful short-term outcomes in young patients requiring cardiac pacing.
Area of Science:
- Cardiology
- Pediatric Electrophysiology
- Medical Device Technology
Background:
- Pediatric pacemaker recipients face lifelong risks from repeated fluoroscopy exposure.
- His bundle pacing presents a potential alternative for improved long-term pacing in pediatric populations.
- Minimizing radiation exposure is crucial for young patients undergoing cardiac procedures.
Purpose of the Study:
- To evaluate the feasibility and short-term efficacy of His bundle pacing guided by 3D electroanatomic mapping in pediatric patients.
- To assess the potential of this technique to reduce fluoroscopy time and radiation dose.
- To determine the safety and effectiveness of His bundle pacing in a pediatric cohort, including those with complex congenital heart defects.
Main Methods:
- Utilized an Octapolar Livewire catheter for His bundle mapping and localization.
- Employed the EnSite Precision 3D mapping system to create patient-specific cardiac geometry.
- Recorded His bundle signal strengths and ventricular activation times using the 3D mapping system.
Main Results:
- Successfully implemented selective His bundle pacing in three pediatric patients (ages 10-15).
- Achieved significantly reduced fluoroscopy times (6.4-9.9 minutes) and radiation doses (5.64-10.18 mGy).
- Demonstrated stable lead performance (thresholds, impedances, sensing) at 4-5 months follow-up.
Conclusions:
- His bundle pacing guided by 3D mapping is a successful and safe short-term strategy for pediatric patients.
- This approach is effective even in pediatric patients with altered His bundle physiology, such as those with atrioventricular canal defects.
- The technique offers a valuable method to minimize fluoroscopy exposure in young individuals requiring pacemakers.