Penile plethysmography on impotent men using vacuum constrictor devices.

J L Marmar1, T J DeBenedictis, D E Praiss

  • 1Division of Urology, Cooper Hospital/University Medical Center, Robert Wood Johnson Medical School, Camden, New Jersey.

Urology
|September 1, 1988
PubMed
Summary

Vacuum constrictor devices maintain penile blood flow during use. Penile plethysmography suggests constrictor rings are safe for erectile dysfunction patients, aiding device selection.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

A minimally invasive vasectomy with the no suture, inline method for vas occlusion.

International journal of fertility and women's medicine·2001
Same author

Evaluation of prostate volume by transrectal ultrasonography for use in a brachytherapy program.

Urology·2001
Same author

The pathophysiology of varicoceles in the light of current molecular and genetic information.

Human reproduction update·2001
Same author

The diagnosis and treatment of male infertility in the new millennium.

International journal of fertility and women's medicine·2001
Same author

A modified acrosome induction test.

Archives of andrology·2000
Same author

Modified vasoepididymostomy with simultaneous double needle placement, tubulotomy and tubular invagination.

The Journal of urology·2000

Area of Science:

  • Urology
  • Vascular Physiology

Background:

  • Vacuum constrictor devices (VCDs) are noninvasive treatments for erectile dysfunction.
  • Clinical use of VCDs is limited by a lack of data on their effects on penile blood flow.

Purpose of the Study:

  • To evaluate the impact of VCD constrictor rings on penile blood flow using penile plethysmography.
  • To assess the safety and efficacy of VCDs in patients with varying penile brachial index (PBI) values.

Main Methods:

  • Penile plethysmography was used to measure penile blood flow in 51 men.
  • Measurements were taken before, during, and after the application of a VCD constrictor ring.

Main Results:

  • Constriction caused a 70-75% decline in pulse-volume curve amplitude, but continuous blood flow was maintained.

Related Experiment Videos

  • Penile blood flow returned to baseline within 60 seconds of ring removal in all patients.
  • Post-constriction hyperemia was observed in 33 patients, suggesting transient blood flow increase.
  • Conclusions:

    • VCD constrictor rings appear safe for patients, maintaining adequate penile blood flow.
    • Penile plethysmography is a valuable tool for patient selection and ensuring proper constrictor ring fit.
    • Regular follow-up assessments are recommended for men using VCDs.