Evaluation of Microwave Ablation Efficacy by Strain Elastography and Shear Wave Elastography in ex Vivo Porcine Liver
Related Experiment Video
Updated: Nov 1, 2025

Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Chunyue Luo1, Tao Li1, Zuojia Li2
1Department of Ultrasound Imaging, Affiliated Cancer Hospital & Institute of Guangzhou Medical University, Guangzhou, Guangdong Province, China.
Abstract:
The aim of this study was to evaluate the efficacy of microwave ablation by ultrasound (US), strain elastography (SE) and shear-wave elastography (SWE). An ex vivo model of porcine liver was adopted. According to ablation power and duration, 30 samples were divided into three groups: group 1 (45 W, 30 s), group 2 (45 W, 15 s) and group 3 (30 W, 30 s). US was used to measure the largest transverse diameter (D1), vertical diameter (D2) and anteroposterior diameter (D3) of the ablated area. SE was used to measure the largest transverse diameter (SEL1), vertical diameter (SEL2) and anteroposterior diameter (SEL3). The actual size of the ablated area was measured as the largest transverse diameter (L1), vertical diameter (L2) and anteroposterior diameter (L3). SWE values and temperatures were measured in the central lesion (region a), marginal area (region b) and unablated area (region c). At 1 h post-ablation, the values measured by US (D1, D2, D3) were all significantly smaller than the ablated area (L1, L2, L3) in all three groups. Except for SEL2 in group 1, there was no significant difference in the results between SEL and L among the three groups. All SWE results were significantly higher post-ablation than pre-ablation in the central lesion (region a) and marginal area (region b, all p values <0.05). In regions a, b and c, the temperatures measured immediately and 5 min post-ablation were all higher than that measured pre-ablation. These results suggest that SE and SWE can be used to evaluate the ablation efficacy of liver tissue.
More Related Videos
10:10Measurement of Liver Stiffness Using Atomic Force Microscopy Coupled with Polarization Microscopy
Published on: July 20, 2022
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019