Related Experiment Video
Updated: Nov 1, 2025

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Change in the local impedance and electrograms recorded by a micro-electrode tip catheter during initial atrial
Kenji Hashimoto1, Ippei Tsuzuki1, Yuta Seki1
1Department of Cardiology Keio University School of Medicine Tokyo Japan.
Background:
A novel measurement of the local impedance (LI) and electrograms recorded from micro-electrodes on catheter tip has been developed. However, the data during pulmonary vein (PV) ablation is not sufficient. We aimed to investigate the utility of this measurement during initial atrial fibrillation (AF) ablation.
Methods:
We investigated 111 representative radiofrequency applications in 7 AF patients without a history of prior ablation (6 males, age 68 [65-72] years, 2 persistent AF). The ablation strategy was PV isolation for paroxysmal AF and single ring box isolation for persistent AF, using MiFi catheter. The correlation of the generator impedance (GI) drop and LI drop after radiofrequency applications and the predictive value of the initial LI elevation before radiofrequency applications for LI drop were analyzed. Also, the LI and GI drop were investigated according to the location of RF applications.
Results:
The LI drop was higher than GI drop (23.7 [16.4-35.7] and 9.0 [6.0-12.0]; P < .01). There were correlations between the initial LI elevation and LI drop (R 2 = 0.466, P < .01) and between the LI and GI drop (R 2 = 0.263, P < .01). The LI drops significantly differed according to the different anatomical localizations by the Kruskal-Wallis test, although the GI drops did not differ (P < .01 and P = .49, respectively).
Conclusion:
LI drop was associated with initial LI elevation and was larger than GI drop. LI drop was different according to locations, although GI drop was not. These findings might indicate that LI drop would be a more sensitive marker for lesion formation than GI drop.