Twenty-Four-Hour Ambulatory Blood Pressure Measurement Using a Novel Noninvasive, Cuffless, Wireless Device
Dean Nachman1,2, Adi Gilan1, Nir Goldstein3
1Department of Military Medicine, Institute for Research in Military Medicine, Faculty of Medicine, The Hebrew University of Jerusalem and the Israel Defense Force Medical Corps, Jerusalem, Israel.
Background:
Ambulatory blood pressure monitoring (ABPM) using cuff-based devices is used for diagnosis and treatment of hypertension. Technical limitations, low compliance, and complex procedures limit their use. The aim of the present study was to test the accuracy of a new photoplethysmography-based, wearable device (Wrist-monitor) as compared with the standard cuff-based ABPM device.
Methods:
Twenty-four-hour (24H) ABPM was performed in parallel for both devices on volunteers aged 18-65 years, while documenting their daily activities. Level of comfort and activity disturbance of both devices were recorded. Linear regression and Bland-Altman were used to evaluate the agreement between devices. Receiver operating characteristic (ROC) curve analysis was used to classify hypertension based on the average Wrist-monitor measurements as compared with a cuff-based ABPM device.
Results:
The study included 28 subjects (18 men) mean age 41.5 ± 16.2 years. Bland-Altman analysis resulted in 24H bias of -1.1 mm Hg for both diastolic blood pressure (DBP) and systolic blood pressure (SBP). Mean daytime bias was -1.9 mm Hg for DBP and SBP, while nighttime bias was smaller (0.7 and 0.4 mm Hg for DBP and SBP, respectively). ROC curve analysis yielded a mean area under the curve (AUC) of 1 for SBP and 24H blood pressure measurements. AUCs of 0.994 and 0.955 were found for the daytime DBP and night DBP, respectively. 24H ABPM with the Wrist-monitor caused significantly less inconvenience compared with the cuff-based device (P < 0.001).
Conclusions:
The cuffless device provides comparable measurements to those obtained with the currently used cuff-based ABPM device, with significantly less inconvenience to the subject.
Clinical Trials Registration:
Trial Number NCT03810586.
More Related Videos
06:51Development of an Algorithm to Perform a Comprehensive Study of Autonomic Dysreflexia in Animals with High Spinal Cord Injury Using a Telemetry Device
Published on: July 29, 2016
08:54Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure