Interference of eltrombopag with bilirubin measurements on the Vitros 5600 analyzer
Izmarie Poventud-Fuentes1,2, Emily Garnett1,2, Jenny Despotovic3
1Department of Pathology and Immunology, Baylor College of Medicine, Houston, TX, USA.
Background:
Eltrombopag is a thrombopoietin-receptor agonist used to restore platelet count to hemostatic levels in chronic immune thrombocytopenia. The drug has shown to have hepatobiliary adverse effects, but also positive interference with the analytical measurement of bilirubin. Understanding the degree of interference of this drug with bilirubin testing becomes relevant in the clinical management of these patients.
Methods:
Eltrombopag at concentrations ranging from 10 to 150 µg/ml was spiked into plasma samples with different baseline concentrations of bilirubin. Total bilirubin, conjugated, and unconjugated bilirubin were measured for each sample using VITROS TBILI and BuBc slides on the Vitros 5600 automated chemistry platform, and interference was assessed.
Results:
Plasma samples spiked with eltrombopag yielded falsely elevated bilirubin measurements compared to baseline, with the degree of elevation increasing with greater concentrations of eltrombopag. Bilirubin values were increased relative to baseline across all groups, except in conjugated bilirubin measurements in samples with low baseline conjugated bilirubin. For samples with low total bilirubin at baseline, >100 µg/ml of eltrombopag resulted in an error of >+0.6 mg/dl on the measured total bilirubin. For samples with low unconjugated bilirubin at baseline, the error for the same concentrations was >+0.7 mg/dl.
Conclusion:
Our results show that, at supra-physiologically high concentrations, eltrombopag can positively interfere with bilirubin measurements on Vitros 5600 platform.
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...