Minimisation of slab-selective radiofrequency excitation pulse durations constrained by an acceptable aliasing
Merlin J Fair1, Peter D Gatehouse1, David N Firmin1
1CMR Unit, Royal Brompton Hospital, London, UK; National Heart and Lung Institute, Imperial College London, London, UK.
Abstract:
A technical note is presented on the slab-direction aliasing of 3D imaging, introducing a simple methodology for determining the minimised duration of low flip-angle sinc radiofrequency (RF) excitation pulses, with respect to a required slab profile accuracy. The various interdependent factors affected in modifying an RF pulse duration are considered and analysed in the context of a new metric for quantifying the levels of permitted slab-aliasing. A general framework is presented for the selection of standard sinc RF excitation pulses with system-minimised durations, as well as their analysis and validation, and a demonstration of this methodology is performed for an example requirement and scanner. This methodology enables implementation of standard (vendor-generated) RF pulses with minimised duration for a required application, with high confidence in their operational reliability. Parts of such a methodology may also in theory be extended to more advanced RF pulse designs.
Related Concept Videos
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
Bandpass Sampling
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
Upsampling
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Linear Approximation in Frequency Domain
In contrast, nonlinear systems do not inherently possess these properties. However, for small deviations around an operating point, a nonlinear system can often be approximated as linear....