Failure of Cleaning Verification in Pharmaceutical Industry Due to Uncleanliness of Stainless Steel Surface
Published on: August 11, 2017
Drug-drug cross contamination in the Swisslog fully automated medication handling system
Anne Skyggedal1, Flemming Nielsen2, Troels K Bergmann3
1Hospital Pharmacy, Hospital of South West Denmark, Esbjerg, Denmark.
Objective:
The risk of drug-drug cross contamination in drug dispensing robots in hospital pharmacies causes cumbersome restraints to be put on the production of the robot for example by scheduling high-risk drugs to be dispensed at the end of the day. However, we were unable to find published data on the matter, and therefore performed a worst-case scenario study to assess the magnitude of the problem.
Methods:
We measured dexamethasone residue left on the suction cup after the production of 100 and 400 dexamethasone tablets, and after 20 paracetamol tablets used as a negative control.
Results:
We found that 32.9 µg and 49.5 µg of dexamethasone had been transferred to the suction cup in the two experiments. This is approximately 1 per mille of the dexamethasone content in a 40 mg tablet.
Conclusion:
We conclude that uncoated dexamethasone does shed measurable residue in the robot. It remains unknown to what extent this residue contaminates the subsequent production.
More Related Videos
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018
09:07Author Spotlight: Accelerating Diagnostic Accuracy with Direct Identification of Gram-Negatives from Blood Culture Bottles
Published on: May 24, 2024
Related Concept Videos
Oral Drug Delivery Systems: Continuous-Release Systems
Pharmaceutical Poisoning: Potential Scenarios
Automated Microbial Diagnostics