Pseudo-elevation of conduction system pacing threshold through parallel connection of an intracardiac electrogram
Jun Yokota1, Katsuhito Fujiu2,3, Tsukasa Oshima2
1Department of Medical Engineering, The University of Tokyo Hospital, Tokyo, Japan.
Journal of Cardiovascular Electrophysiology
|June 24, 2021
Summary
Connecting an electrophysiology (EP) system in parallel with pacing equipment can unexpectedly increase pacing thresholds during conduction system pacing (CSP). Disabling the EP system's pacing function resolves this issue, preventing false CSP failures.
Area of Science:
- Cardiovascular Electrophysiology
- Medical Device Engineering
Background:
- Parallel connection of electrophysiology (EP) systems to conduction system pacing (CSP) equipment is common for monitoring and evaluation.
- Unexpected increases in pacing thresholds have been observed under specific parallel connection conditions.
Purpose of the Study:
- To investigate the mechanism behind unexpected pacing threshold exacerbation when an EP system is connected in parallel with CSP equipment.
- To evaluate the impact of EP system settings on pacing output and intracardiac electrograms.
Main Methods:
- An ex vivo model simulating cardiac pacing and electrogram monitoring was created.
- An oscilloscope, pacing system analyzer (PSA), EP system, and simulated heart were used.
- Electrical output discrepancies were measured under various equipment combinations and conditions.
Main Results:
- Parallel connection of the EP system reduced the electrical output from the PSA, as measured at the simulated heart.
- Adverse effects were pronounced with the RMC-5000 EP system when its pacing function was active.
- Disabling the EP system's pacing function completely resolved the observed reduction in electrical output.
Conclusions:
- The EP system, when connected in parallel with a PSA for CSP, has the potential to increase pacing thresholds.
- This phenomenon may lead to pseudo-failures in CSP due to artificially high pacing thresholds.
- It is recommended to turn off the pacing function of the RMC-5000 EP system when used in parallel with a PSA for CSP testing.
Related Concept Videos
Electrocardiogram
4.3K
An electrocardiogram (ECG or EKG) is a critical diagnostic tool that records the electrical signals produced by the heart during each heartbeat. This recording is achieved through electrodes placed strategically on the arms, legs, and chest. The electrocardiograph amplifies these signals and produces 12 distinct tracings, offering a comprehensive understanding of the heart's electrical activity.
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
4.3K
Electrocardiogram Fundamentals
1.0K
Introduction
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
1.0K
Pulse rhythm
1.0K
Pulse rhythm refers to the pattern of pulsations within specific intervals, offering valuable insights into the regularity or irregularity of the heart's beats as observed through the pattern of pulsation within specific intervals. A regular pulse exhibits a consistent heart rate with uniform waveforms and pulsation force, variations of which can be classified as normal, weak, or bounding.
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
1.0K
Electrophysiology of Normal Cardiac Rhythm
7.8K
The normal cardiac rhythm is a synchronized electrical activity that facilitates the regular and coordinated contraction of the heart muscle. This process is essential for efficient blood circulation throughout the body. The fundamental elements involved in establishing and maintaining this rhythm include the unique electrical properties of cardiac muscle cells, the sinoatrial (SA) node's pacemaker function, the specialized conducting system, and the ionic mechanisms underlying each phase...
7.8K