Simultaneous determination of monocrotaline and its N-oxide metabolite in rat plasma using LC-MS/MS: Application to a
Feifei Lin1,2, Anni Pan2, Yang Ye1,2
1School of Chinese Materia Medica, Nanjing University of Chinese Medicine, Nanjing, China.
Abstract:
Monocrotaline (MCT) is a pyrrolizidine alkaloid that can induce hepatic sinusoidal damage, pulmonary hypertension, renal toxicity, and heart disease. Monocrotaline N-oxide (MNO), the primary metabolite of MCT, is less toxic; however, it can convert back to MCT to exhibit its toxicity. This study developed and validated a rapid and sensitive LC-MS/MS method for the simultaneous determination of MCT and monocrotaline N-oxide in rat plasma. The method has a linearity over the concentration range of 1-2000 ng/mL with correlation coefficients (r) >0.997 for each analyte. The results of selectivity, matrix effect, accuracy and precision, and recovery were all within the acceptance criteria. The validated method has been successfully applied to study pharmacokinetic behaviors and bioavailability of MCT in rats. MCT was rapidly absorbed (Tmax : 0.400 ± 0.149 h) after oral administration, and the absolute bioavailability of MCT was 78.2%.