Pediatric reference interval verification for endocrine and fertility hormone assays on the Abbott Alinity system
Mary Kathryn Bohn1,2, Siobhan Wilson1,2, Alexandra Hall1
1CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, ON, Canada.
Objectives:
The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) has developed an extensive database of reference intervals (RIs) for several biomarkers on various analytical systems. In this study, pediatric RIs were verified for key immunoassays on the Abbott Alinity system based on the analysis of healthy children samples and comparison to comprehensive RIs previously established for Abbott ARCHITECT assays.
Methods:
Analytical performance of Alinity immunoassays was first assessed. Subsequently, 100 serum samples from healthy children recruited with informed consent were analyzed for 16 Alinity immunoassays. The percentage of test results falling within published CALIPER ARCHITECT reference and confidence limits was determined. If ≥ 90% of test results fell within the confidence limits, they were considered verified based on CLSI guidelines. If <90% of test results fell within the confidence limits, additional samples were analyzed and new Alinity RIs were established.
Results:
Of the 16 immunoassays assessed, 13 met the criteria for verification with test results from ≥ 90% of healthy serum samples falling within the published ARCHITECT confidence limits. New CALIPER RIs were established for free thyroxine and prolactin on the Alinity system. Estradiol required special considerations in early life.
Conclusions:
Our data demonstrate excellent concordance between ARCHITECT and Alinity immunoassays, as well as the robustness of previously established CALIPER RIs for most immunoassays, eliminating the need for de novo RI studies for most parameters. Availability of pediatric RIs for immunoassays on the Alinity system will assist clinical laboratories using this new platform and contribute to improved clinical decision-making.