A Randomized Comparison of Delivered Energy in Cardioversion of Atrial Fibrillation: Biphasic Truncated Exponential
Elina Trendafilova1, Elena Dimitrova1, Jean-Philippe Didon2
1Intensive Cardiology Care Unit, Cardiology Clinic, National Cardiology Hospital, 65 Konyovitza Str., 1309 Sofia, Bulgaria.
Abstract:
A few randomized trials have compared impedance-compensated biphasic defibrillators in clinical use. We aim to compare pulsed biphasic (PB) and biphasic truncated exponential (BTE) waveforms in a non-inferiority cardioversion (CVS) study. This was a prospective monocentric randomized clinical trial. Eligible patients admitted for elective CVS of atrial fibrillation (AF) between February 2019 and March 2020 were alternately randomized to treatment with either a PB defibrillator (DEFIGARD TOUCH7, Schiller Médical, Wissembourg, France) or a BTE high-energy (BTE-HE) defibrillator (LIFEPAK15, Physio-Control Inc., Redmond, WA, USA). Fixed-energy protocol (200-200-200 J) was administered. CVS success was accepted if sinus rhythm was restored at 1 min post-shock. The study design considered non-inferiority testing of the primary outcome: cumulative delivered energy (CDE). Seventy-three out of 78 randomized patients received allocated intervention: 38 BTE-HE (52%), 35 PB (48%). Baseline characteristics were well-balanced between groups (p > 0.05). Both waveforms had similar CDE (mean ± standard deviation, 95% confidence interval): BTE-HE (253.9 ± 120.2 J, 214-293 J) vs. PB (226.0 ± 109.8 J, 188-264 J), p = 0.31. Indeed, effective PB shocks delivered significantly lower energies by mean of 25.6 J (95% CI 24-27.1 J, p < 0.001). Success rates were similar (BTE-HE vs. PB): 1 min first-shock (84.2% vs. 82.9%), 1 min CVS (97.4% vs. 94.3%), 2 h CVS (94.7% vs. 94.3%), 24 h CVS (92.1% vs. 94.3%), p > 0.05. Safety analysis did not find CVS hazards, reporting insignificant changes of myocardial-specific biomarkers, transient and rare ST-segment deviations, and no case of harmful tachyarrhythmias and apnea. Cardioversion of AF with fixed-energy protocol 200-200-200 J was highly efficient and safe for both PB and BTE-HE waveforms. These similar performances were achieved despite differences in the waveforms' technical design, associated with significantly lower delivered energy for the effective PB shocks. Clinical Trial Registration: Registration number: NCT04032678, trial register: ClinicalTrials.gov.
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
Dysrhythmias III: Characteristics of Dysrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials
Voltammetric Techniques: Pulse Voltammetry