Bilateral metallic hip implants: Are avoidance sectors necessary for pelvic VMAT treatments?
Nikolaos Koutsouvelis1, Giovanna Dipasquale1, Michel Rouzaud1
1Department of Radiation Oncology, Geneva University Hospital, Geneva, Switzerland.
Purpose:
Metallic hip implants (MHI) are common in elderly patients. For pelvic cancers radiotherapy, conventional approaches consist of MHI avoidance during treatment planning, which leads, especially in case of bilateral MHI, to a decreased quality or increased complexity of the treatment plan. The aim of this study is to investigate the necessity of using avoidance sectors (AvSe) using a 2-arcs coplanar pelvic volumetric modulated arc-therapy (VMAT) planning.
Methods:
We evaluated: (1) The dose calculation error of a static 6MV open beam traversing a MHI; (2) The magnitude of an error's decrease within the planning target volume (PTV) for a 360° VMAT treatment without AvSe as compared to the static open beam; (3) The dosimetric influence of MHI misalignment generated by patient's repositioning rolls during image-guided radiotherapy (IGRT).
Results:
(1) In the static 6MV beam configuration, for distances between 0.5cm and 6cm from the MHI, the median (maximum, number of points) dose calculation error was -1.55% (-2.5%, 11); (2) Compared to the static open beam, in the 360° VMAT treatment without AvSe a simulated error was decreased by a factor of 4.4/2.4 (median/minimum); (3) MHI anterior-posterior misalignment exceeding 0.6cm, resulted in error at PTV surface of >2%.
Conclusions:
A standard 2 coplanar arcs 360° VMAT treatment, with dedicated artifact reduction algorithms applied, decreased the error of static beam traversing MHI, in patients presenting a bilateral MHI and might be used to treat the pelvic region without MHI avoidance.