Comparison of Ischial Containment and Subischial Sockets on Comfort, Function, Quality of Life, and Satisfaction With
Stefania Fatone1, Ryan Caldwell2, John Angelico3
1Department of Physical Medicine and Rehabilitation, Northwestern University, Feinberg School of Medicine, Chicago, IL.
Objective:
To compare comfort and functional performance of the Northwestern University Flexible Subischial Vacuum (NU-FlexSIV) Socket with the ischial containment (IC) socket in persons with unilateral transfemoral amputation.
Design:
Randomized crossover trial with two 7-week periods.
Setting:
Private prosthetic clinics and university research laboratory.
Participants:
A total of 30 enrolled (N=30); 25 participants completed the study with full (n=18) or partial data (n=7).
Interventions:
Two custom-fabricated sockets (IC and NU-FlexSIV), worn full-time for 7 weeks, with testing at 1, 4, and 7 weeks after socket delivery.
Main Outcome Measures:
The primary outcome was change in Socket Comfort Score (SCS) at 7 weeks. Secondary outcomes at 7 weeks included the Orthotic and Prosthetic Users' Survey (OPUS) to assess lower extremity functional status, health-related quality of life, and satisfaction with device, as well as the 5-Times Rapid Sit-to-Stand Test, Four Square Step Test, and T-Test of Agility to assess functional performance.
Results:
At 7 weeks, the mean SCS for IC (7.0±1.7) and NU-FlexSIV (8.4±1.1) Sockets were significantly different (P<.001; 95% confidence interval, 0.8-2.3). Results from a linear mixed-effects model, accounting for data from all time points, indicated that the SCS was 1.7 (SE=0.45) points higher for the NU-FlexSIV Socket (P<.001). For the secondary outcomes, only OPUS satisfaction with device was significantly better in the NU-FlexSIV Socket after accounting for all data points.
Conclusions:
The results suggest that after 7 weeks' accommodation, the NU-FlexSIV Socket was more comfortable and led to greater satisfaction with device than the IC socket in persons with unilateral transfemoral amputation and K3/K4 mobility. Other patient-reported outcomes and function were no different between sockets.