Related Experiment Video
Updated: Oct 30, 2025

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Radioactivity measurement of 85Kr by using anti-cosmic HPGe γ spectrometer
Qi Li1, Shujing Liu1, Shilian Wang1
1CTBT Beijing National Data Centre and Beijing Radionuclide Laboratory, Beijing, China.
Abstract:
In order to improve the detection sensitivity of 85Kr, an anti-cosmic HPGe γ spectrometer was established and a specific Marinelli beaker was designed for 85Kr measurement. Comparing to the non-anticoincidence γ spectrum, the integral background counts rate range from 20 to 2400 keV was reduced by a factor of 5.17 using anticoincidence shielding. The minimum detectable activity of 85Kr was 10.5 Bq within 24 h measurement in standard mode and that of 3.99 Bq in anticoincidence mode. The activity concentration of atmospheric 85Kr was preliminarily measured to be an average of 1.30 Bq/m3 in Beijing and corresponding minimum detectable concentration was 0.2 Bq/m3.