Related Experiment Video
Updated: Oct 30, 2025

Comprehensive Evaluation of the Effectiveness and Safety of Placenta-Targeted Drug Delivery Using Three Complementary Methods
Published on: September 10, 2018
Bioanalytical RP-HPLC method for the determination of ponazuril in plasma
Sherry Cox1, Lainey Harvill1, Sarah Singleton1
1Department of Biomedical and Diagnostic Sciences, College of Veterinary Medicine, The University of Tennessee, Knoxville, TN, USA.
Abstract:
The goal of this investigation was to establish a reliable technique for the quantitation of ponazuril in limited sample volumes. Samples were extracted with chloroform and separation was achieved with a Symmetry RP18 column. Ultraviolet absorption was measured at 254 nm. A combination of 0.1% formic acid and acetonitrile (50:50) was used as the mobile phase. The calibration curve was linear from 0.1-25 μg/mL, with a lower limit of quantification of 0.1 μg/mL with a 100 μL sample. The precision and accuracy were well within the range set by the Food and Drug Administration and the recovery was over 95%. This technique was used to analyze ponazuril samples and found to be appropriate for pharmacokinetic studies.