Validation technique and improvements introduced in a new dedicated brain positron emission tomograph (CareMiBrain)
María Nieves Cabrera-Martín1, Gabriel González-Pavón2, Miguel Sanchís Hernández2
1Servicio de Medicina Nuclear, Hospital Clínico San Carlos, Instituto de Investigación Sanitaria San Carlos (IdISSC), Universidad Complutense, Madrid, Spain.
Abstract:
The goal of developing a PET dedicated to the brain (CareMiBrain) has evolved from its initial approach to diagnosis and monitoring of dementias, to the more ambitious of creating a revolutionary clinical pathway for the knowledge and personalized treatment of multiple neurological diseases. The main innovative feature of CareMiBrain is the use of detectors with continuous crystals, which allow a high resolution determination of the depth of annihilation photons interaction within the thickness of the scintillation crystal. The technical validation phase of the equipment consisted of a pilot, prospective and observational study whose objective was to obtain the first images (40 patients), analyze them and make adjustments in the acquisition, reconstruction and correction parameters, comparing the image quality of the CareMiBrain equipment with that of the whole-body PET/CT. Thanks to the team meetings and the joint analysis of the images, it was possible to detect its weak points and some of its causes. The calibration, acquisition and processing processes, as well as the reconstruction, were optimized, the number of iterations was set to achieve the best signal-to-noise ratio, the random correction was optimized and a post-processing algorithm was included in the reconstruction algorithm. The main technical improvements implemented in this phase of technical validation carried out through collaboration of the Services of Nuclear Medicine and Neurology of the Hospital Clínico San Carlos with the Spanish company Oncovision will be exposed in a project financed with funds from the European Union (Horizon 2020 innovation program, 713323).
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET