Related Experiment Video
Updated: Oct 30, 2025

Conformable Wearable Electrodes: From Fabrication to Electrophysiological Assessment
Published on: July 22, 2022
Noncontact ECG Monitoring by Capacitive Coupling of Textiles in a Chair
Po-Cheng Su1, Ya-Hsin Hsueh1, Ming-Ta Ke2
1Department of Electronic Engineering, National Yunlin University of Science and Technology, Douliu 64002, Taiwan.
Abstract:
Some patients are uncomfortable with being wired to a device to have their heart activity measured. Accordingly, this study adopts a noncontact electrocardiogram (ECG) measurement system using coupled capacitance in a conductive textile. The textiles can be placed on a chair and are able to record some of the patient's heart data. Height and distance between the conductive textile electrodes were influential when trying to obtain an optimal ECG signal. A soft and highly conductive textile was used as the electrode, and clothing was regarded as capacitance insulation. The conductive textile and body were treated as the two electrode plates. This study found that placing the two conductive textiles at the same height provided better data than different heights. The system also enabled identifying the P, Q, R, S, and T waves of the ECG signal and eliminated unnecessary noise successfully.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...