Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall
Jos Thannhauser1, Joris Nas1, Priya Vart2
1Department of Cardiology, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
Aim:
In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to cases with a high AMSA. Appreciating the variability of in-field pad placement, we aimed to assess the impact of recording direction on AMSA-values, and the inherent defibrillation advice.
Methods:
Prospective VF-waveform study on 12-lead surface electrocardiograms (ECGs) obtained during defibrillation testing in ICD-recipients (2010-2017). AMSA-values (mVHz) of simultaneous VF-recordings were calculated and compared between all limb leads, with lead II as reference (proxy for in-field pad position). AMSA-differences between leads I and II were quantified using Bland-Altman analysis. Moreover, we investigated differences between these adjacent leads regarding classification into high (≥15.5), intermediate (6.5-15.5) or low (≤6.5) AMSA-values.
Results:
In this cohort (n = 243), AMSA-values in lead II (10.2 ± 4.8) differed significantly from the other limb leads (I: 8.0 ± 3.4; III: 12.9 ± 5.6, both p < 0.001). The AMSA-value in lead I was, on average, 2.24 ± 4.3 lower than in lead II. Of the subjects with high AMSA-values in lead II, only 15% were classified as high if based on assessments of lead I. For intermediate and low AMSA-values, concordances were 66% and 72% respectively.
Conclusions:
ECG-recording direction markedly affects the result of VF-waveform analysis, with 20-30% lower AMSA-values in lead I than in lead II. Our data suggest that electrode positioning may significantly impact shock guidance by 'smart defibrillators', especially affecting the advice for prompt defibrillation.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials