Validation of a flash glucose monitoring system in outpatient diabetic cats
Emily K Shea1, Rebecka S Hess1
1Department of Clinical Sciences and Advanced Medicine, School of Veterinary Medicine, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
Background:
Interstitial glucose (IG) concentration measurement using a flash glucose monitoring system (FGMS) is a noninvasive, affordable, and informative method to regulate patients with diabetes mellitus (DM) but has not been fully validated in outpatient cats with DM.
Objectives:
To further validate the FreeStyle Libre FGMS in outpatient diabetic cats.
Animals:
Eight client-owned cats with DM.
Methods:
Prospective observational validation study. Tissue glue was used to attach the sensor to the cat. Lin's concordance correlation coefficient (ρc ) was used to compare IG concentrations measured by the FGMS to blood glucose concentrations measured using an automated biochemistry analyzer (ABA) and point-of-care glucometer (POCG).
Results:
Data from 15 sensor placements in 8 cats were analyzed. Paired IG and ABA glucose concentrations (139 samples) had excellent correlation (ρc = 0.96) as did IG and POCG glucose concentrations (142 samples, ρc = 0.92). Sensor failure or displacement were recorded for 12/15 (80%) sensor placements. Median time of sensor activity was 7 days (range, 2-13 days).
Conclusions And Clinical Importance:
In outpatient cats with DM, the FGMS-measured IG concentration correlated well with ABA-measured blood glucose concentration, but a high rate of sensor failures was observed.