Related Experiment Video
Updated: Oct 29, 2025

Plasma Polishing as a New Polishing Option to Reduce the Surface Roughness of Porous Titanium Alloy for 3D Printing
Published on: April 28, 2023
Femtosecond laser-induced nanoporous layer for enhanced osteogenesis of titanium implants
Hao Yao1, Xianrui Zou1, Shijian Zheng1
1Tianjin Key Laboratory of Materials Laminating Fabrication and Interface Control Technology, School of Materials Science and Engineering, Hebei University of Technology, Tianjin 300130, China.
Abstract:
The osteogenic activity of medical metal can be improved by lowering its surface stiffness and elastic modulus. However, it is very difficult to directly reduce the elastic modulus of medical metal surfaces. In this paper, with selected parameters, the titanium surface was treated via femtosecond laser irradiation. Micro indentation revealed that the femtosecond laser ablation can effectively reduce the surface Young's modulus and Vickers hardness of titanium. Besides, In order to explain the mechanical properties of degradation of titanium surface, Large-scale Atomic/Molecular Massively Parallel Simulator (LAMMPS) was used to simulate the process of laser ablation process of titanium surface, and it was found that after the ablation of titanium surface, voids were produced in the subsurface layer. The simulation showed that the voids are formed by the cavitation of metastable liquid induced by high tensile stress and high temperature during femtosecond laser irradiation. Subsurface voids with a thickness of about 40 nm were observed under the oxide layer in the experiment. Cell experiments showed that the surface with low Young's modulus was more conducive to cell proliferation and osteogenic differentiation.