Impact of prostate focused alignment on planned pelvic lymph node dose
Joshua Kilian-Meneghin1, Tianjun Ma1, Lalith Kumaraswamy1
1Roswell Park Comprehensive Cancer Center, Buffalo, NY, USA.
Journal of Applied Clinical Medical Physics
|July 7, 2021
Summary
Prostate cancer patients receiving radiation therapy may experience reduced pelvic lymph node coverage due to motion discrepancies. Accounting for lymph node motion during treatment alignment is crucial for effective prostate cancer radiotherapy.
Area of Science:
- Radiation Oncology
- Medical Physics
Background:
- Prostate cancer patients with positive lymph node margins undergo radiation therapy targeting the prostate, seminal vesicles, and lymph nodes.
- External beam therapy (EBT) for these patients involves a 45 Gy dose to the planning target volume (PTV).
- Prostate localization uses fiducial markers and portal imaging, but pelvic lymph nodes are fixed to bony anatomy, potentially causing coverage discrepancies.
Purpose of the Study:
- To evaluate the difference between planned and delivered pelvic lymph node coverage in prostate cancer patients undergoing EBT.
- To determine if pelvic lymph node motion significantly impacts radiation dose delivery.
Main Methods:
- Prostate and pelvic lymph node motions were tracked for 19 patients using kV-images and bony anatomy registration.
- Deviations were calculated and input as field shifts in Eclipse to simulate planned vs. delivered coverage.
- Dose coverage was assessed using V(100) criteria for lymph node clinical target volume (LN CTV) and lymph node planning target volume (LN PTV).
Main Results:
- A significant difference (P=0.005) was observed in planned vs. delivered dose coverage for the LN PTV, averaging 3.5%.
- 26% of patients received less than 95% dose coverage for the LN PTV based on bony anatomy registration.
- No significant difference in coverage was found for the LN CTV (P=0.53).
Conclusions:
- A significant discrepancy exists between planned and delivered radiation dose coverage for the pelvic lymph node planning target volume (LN PTV).
- Pelvic lymph node motion should be considered alongside prostate motion during patient alignment for EBT.
- Accurate patient setup is critical to ensure adequate lymph node coverage in prostate cancer radiotherapy.