Relation Between Insertion Torque and Implant Stability Quotient: A Clinical Study
Related Experiment Video
Updated: Oct 29, 2025

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
João Paulo do Vale Souza1, Clóvis Lamartine de Moraes Melo Neto1, Lucas Tavares Piacenza1
1Department of Dental Materials and Prosthodontics, Araçatuba Dental School, São Paulo State University, Araçatuba, São Paulo, Brazil.
Objectives:
This study aimed to assess the relation between the insertion torque and implant stability quotient (ISQ recorded immediately and 6 months after implant placement).
Materials And Methods:
Twenty-five patients over the age of 18 years were selected for this study. One implant was placed per patient after tooth extraction. All implants had the same size (11.5 × 3.75 mm) and brand (Hexagonal Morse cone, DSP Biomedical). The insertion torque (Ncm) and resonance frequency analysis (ISQ value) (Osstell Mentor) were used to assess the primary stability (on the day of surgery). After 6 months, resonance frequency analysis was used to assess the secondary stability of each implant.
Statistical Analysis:
The insertion torque data were correlated with ISQ measurements by using Pearson's correlation. The significance level was 5%.
Results:
There was a positive correlation between insertion torque and initial ISQ (correlation: 0.457; p = 0.022); however, no correlation was found between insertion torque and final ISQ (p = 0.308).
Conclusion:
The present study demonstrated that there is a positive correlation between the insertion torque and the initial ISQ. Therefore, the higher the insertion torque, the higher the initial ISQ (and vice versa).