A retrospective analysis of discordances between international normalized ratio (INR) self-testing and INR laboratory

Joan Bitan1, Fanny Bajolle2,3, Annie Harroche4

  • 1Hematology Laboratory, Hôpital universitaire Necker-Enfants Malades, AP-HP, Paris, France.

Insights

Discrepancies between home point-of-care (POC) and laboratory international normalized ratio (INR) testing in children are linked to younger age and the use of animal thromboplastin in lab tests. This highlights the need for specific quality control for POC INR devices.

Area of Science:

  • Pediatric Hematology
  • Clinical Diagnostics
  • Point-of-Care Testing

Background:

  • Quality control for home point-of-care (POC) international normalized ratio (INR) devices is a significant concern.
  • Concomitant laboratory and POC INR testing can help address quality control issues.
  • Differences between POC and laboratory INR results may not always indicate POC device failure.

Purpose of the Study:

  • To identify variables associated with significant deviations between POC INR and laboratory INR measurements.
  • To investigate factors influencing the accuracy of home INR monitoring in pediatric patients.

Main Methods:

  • Retrospective cohort study involving children with paired laboratory and POC INR measurements.
  • Analysis of clinical and laboratory variables to determine associations with significant INR deviations.
  • Multivariate analysis to identify independent predictors of INR discrepancies.

Main Results:

  • Significant deviations occurred in 15.3% of INR measurement pairs (30/196) from 124 children.
  • Younger age (OR=0.91, P=.020) and laboratory use of animal thromboplastin (OR=2.81, P=.016) were associated with significant deviations.
  • Less experience with POC INR devices showed a trend towards association with deviations (OR=0.89, P=.098).

Conclusions:

  • While most INR pairs were consistent, deviations highlight the impact of laboratory thromboplastin origin.
  • The findings underscore the necessity for specific quality control procedures for POC INR devices in pediatric care.
  • Further investigation into thromboplastin sources and POC device validation is warranted.
Abstract