MRI of Traumatic Knee Dislocation: A Study to Evaluate Safety and Image Quality for Patients with Knee-Spanning
Lawrence Lo1, Shams Jubouri1, Michael E Mulligan2
1Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, School of Medicine, Baltimore, MD.
Abstract:
This study evaluated safety and image quality of MRI exams performed for patients with traumatic knee dislocations in knee-spanning stabilization devices. It is an IRB-approved retrospective design with waived informed consent that included 63 patients with traumatic knee dislocation. 56 patients had metallic external fixators, and 7 patients had non-metallic knee immobilizers. 7 patients had bilateral dislocations yielding a total of 70 knee MRIs. 1.5 Tesla MRI exams were performed for all patients who were awake and alert at the time of imaging. All knee-spanning external fixators were considered "MR conditional" by the FDA. The electronic medical record was reviewed for notes from the technologist and nursing staff documenting any patient complaints or adverse events during the MRI exam as required by departmental protocol. Qualitative analysis of the six most frequently performed sequences were independently conducted by 2 musculoskeletal radiologists using a 5-point Likert scale. Overall image quality and select time intervals between the two groups were compared using an independent sample t test and the Mann-Whitney U test, respectively. No adverse events were reported for a 40-minute average estimated patient scan time with the stabilization devices in the MR gantry. Mean values of Likert scale scores were generated from two readers' data for comparison between the external fixation and the immobilizer groups. Most knee MRI exams with external fixators were within diagnostic quality despite artifacts (grade 3). MRI exams generally were of higher diagnostic quality in the immobilizer group than the external fixator group (p < 0.05). The external fixator models included DePuy Synthes, Smith and Nephew, Stryker Hoffman III, Zimmer FastFrame, and Zimmer XtraFix. MRI examinations in patients with external fixators for traumatic knee dislocations can be safely performed under certain conditions and provide diagnostic quality images.
More Related Videos
06:27Author Spotlight: Investigating Early Events and Long-Term Effects of ACL Injuries for Osteoarthritis Progression
Published on: September 29, 2023
06:28Anterior Cruciate Ligament Transection and Synovial Fluid Lavage in a Rodent Model to Study Joint Inflammation and Posttraumatic Osteoarthritis
Published on: September 2, 2025
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI