[Assessment of imaging performance of digital breast tomosynthesis based on systematic simulation]
1School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
Objective:
To assess the imaging performance of digital breast tomography (DBT) based on systematic simulation.
Objective:
The raw measurements of physical phantoms at a variety of radiation dose levels and clinical patients at the normal radiation dose were acquired from a clinical DBT system for low-dose simulation and reconstruction using 3 reconstruction algorithms, namely Feldkamp-Davis-Kress (FDK), simultaneous algebraic reconstruction technique (SART) and adaptive steepest-descent projection onto convex sets with total-variation constraint (ASDPOCS-TV) algorithms. The image quality was compared across different radiation dose levels and reconstruction algorithms in terms of signal-to-noise ratio (SNR), peak signal-to-noise ratio (PSNR), noise-power spectrum (NPS), artifact spread function (ASF) and full width at half maximum (FWHM) of ASF indexes.
Objective:
The reliability of low-dose DBT simulation strategy was verified by the experiment. Within a suitable range of dose levels, increasing the doses resulted in reduced high-frequency noise component and significantly increased SNR (P < 0.05). But when the value of exposure was below 40 mAs, the images acquired at different dose levels had similar representation. The performance of the 3 reconstruction algorithms varied for different anatomical structures, and the image quality of ASDPOCS-TV algorithm was generally superior to SART and FDK algorithms with less through-plane artifacts and noise.
Objective:
The quality of DBT images is significantly affected by both radiation dose and reconstruction algorithms. A tradeoff of the parameters, the overall image quality and the clinical needs for diagnostic purposes should be considered to achieve the optimal imaging performance on a given clinical task.
More Related Videos
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
08:32Using Computer-based Image Analysis to Improve Quantification of Lung Metastasis in the 4T1 Breast Cancer Model
Published on: October 2, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography