Effectiveness of consumer-grade contactless vital signs monitors: a systematic review and meta-analysis
Chi Pham1,2, Khashayar Poorzargar1,2, Mahesh Nagappa3
1Department of Anesthesia and Pain Medicine, Toronto Western Hospital, University Health Network, University of Toronto, 399 Bathurst Street, McL2-405, Toronto, ON, M5T 2S8, Canada.
Abstract:
The objective of this systematic review and meta-analysis was to analyze the effectiveness of contactless vital sign monitors that utilize a consumer-friendly camera versus medical grade instruments. A multiple database search was conducted from inception to September 2020. Inclusion criteria were as follows: studies that used a consumer-grade camera (smartphone/webcam) to examine contactless vital signs in adults; evaluated the non-contact device against a reference medical device; and used the participants' face for measurement. Twenty-six studies were included in the review of which 16 were included in Pearson's correlation and 14 studies were included in the Bland-Altman meta-analysis. Twenty-two studies measured heart rate (HR) (92%), three measured blood pressure (BP) (12%), and respiratory rate (RR) (12%). No study examined blood oxygen saturation (SpO2). Most studies had a small sample size (≤ 30 participants) and were performed in a laboratory setting. Our meta-analysis found that consumer-grade contactless vital sign monitors were accurate in comparison to a medical device in measuring HR. Current contactless monitors have limitations such as motion, poor lighting, and lack of automatic face tracking. Currently available consumer-friendly contactless monitors measure HR accurately compared to standard medical devices. More studies are needed to assess the accuracy of contactless BP and RR monitors. Implementation of contactless vital sign monitors for clinical use will require validation in a larger population, in a clinical setting, and expanded to encompass other vital signs including BP, RR, and SpO2.
More Related Videos
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.