Upgrade and absolute calibration of the JET scintillator-based fast-ion loss detector
J F Rivero-Rodriguez1, C Perez Von Thun2, M Garcia-Muñoz3
1Departamento de Ingenieria Mecanica y Fabricacion, Universidad de Sevilla, Camino de los Descubrimientos s/n, 41092 Sevilla, Spain.
Abstract:
The JET FILD is a scintillator-based Fast-ion Loss Detector optimized to measure fusion-born alpha-particle losses. This work covers its upgrade and absolute calibration in preparation for the following JET DT experiments. A fast scintillator material (TG-Green) has been installed in the JET FILD. A heater jacket is installed around the fiber bundle, responsible for transmitting the light from the scintillator plate, to anneal the fiber obscuring due to neutron damage. The JET FILD has been upgraded with a 1 Mpx camera and 2 MHz photomultiplier data acquisition hardware. Full-orbit simulations give an estimate of the shading effects on the scintillator plate of the first wall structures and provide a synthetic signal of the JET FILD. A detector instrument function enables absolute values of fast-ion losses using calibration factors. The calibration factors are made available in a shot-to-shot basis for the characterized species and energies and with corrections for the diagnostic conditions. The fast acquisition system sets the Nyquist frequency (1 MHz) above the typical mode frequencies (≈102 kHz), thus making it possible to identify MHD-induced fast-ion losses.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Mass Analyzers: Common Types
Gas Chromatography: Types of Detectors-II
Atomic Emission Spectroscopy: Instrumentation