Miniaturized USB-powered multi-channel module for gamma spectroscopy and imaging
Luca Buonanno1, Davide Di Vita1, Marco Carminati1
1Politecnico di Milano, Dipartimento di Elettronica, Informazione e Bioingegneria, 20133 Milano (MI), Italy.
Abstract:
LAILA is a miniaturized eight-channel electronic readout system for compact γ-ray detectors, combining high-resolution spectroscopy capability with position sensitivity. Compactness is achieved by the combination of a novel CMOS front-end ASIC (Application-Specific Integrated Circuit) for analog processing of a large signal current from Silicon PhotoMultiplier (SiPM) solid-state photodetectors, with a microcontroller-based data acquisition system. The adoption of automatic gain regulation in the gated-integrator stage of the ASIC offers an 84 dB dynamic range, combining single-photon sensitivity with an extended input photon energy range (20 keV-4 MeV, using 30 μm-cell SiPMs). Using this module with properly merged 144 SiPM pixels coupled to a 3 in.-thick lanthanum bromide scintillation crystal, a 3% energy resolution at 662 keV and 1 cm spatial resolution in the estimation of the interaction coordinates are experimentally demonstrated in this work.
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
UV–Vis Spectrometers