X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Toward quantitative short-scan cone beam CT using shift-invariant filtered-backprojection with equal weighting and
Abstract:
Quantitative short-scan cone beam CT (CBCT) is impeded by streaking and shading artifacts. Streaking artifacts can be caused by approximate handling of data redundancy in short-scan FDK with Parker's weighting, while shading artifacts are caused by scatter and beam hardening effects. In this work, we improve the image quality of short-scan CBCT by removing the streaking artifacts using a previously proposed algorithm in a framework of filtered backprojection with shift-invariant filtering and equal weighting. An efficient image domain shading correction using sparse samples is subsequently applied to further improve the image uniformity. The improved image quality is shown for this approach, both in visual appearance and quantitative measurements of three clinical head scans.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography