Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
Delivered dose with jet and mesh nebulisers during spontaneous breathing, noninvasive ventilation and mechanical
1Dept of Respiratory Care, Texas State University, Round Rock, TX, USA.
Abstract:
What is the delivered dose with jet and mesh nebulisers during spontaneous breathing (SB), noninvasive ventilation (NIV), and mechanical ventilation (MV) using an adult lung model with exhaled humidity (EH)? The delivery of salbutamol sulfate (2.5 mg per 3 mL) with jet (Mistymax10) and mesh nebulisers (Aerogen Solo) was compared during SB, NIV, and MV using breathing parameters (tidal volume 450 mL, respiratory rate 20 breaths per min, inspiratory:expiratory ratio 1:3) with three lung models simulating exhaled humidity. A manikin was attached to a sinusoidal pump via a filter at the bronchi to simulate an adult with SB. A ventilator (V60) was attached via a facemask to a manikin with a filter at the bronchi connected to a test lung to simulate an adult receiving NIV. A ventilator-dependent adult was simulated through a ventilator (Servo-i) operated with a heated humidifier (Fisher & Paykel) attached to an endotracheal tube (ETT) with a heated-wire circuit. The ETT was inserted into a filter (Respirgard II). A heated humidifier was placed between the filter and test lung to simulate exhaled humidity (35±2°C, 100% relative humidity). Nebulisers were placed at the Y-piece of the inspiratory limb during MV and positioned between the facemask and the leak-port during NIV. A mouthpiece was used during SB. The delivered dose was collected in an absolute filter that was attached to the bronchi of the mannequin during each aerosol treatment and measured with spectrophotometry. Drug delivery during MV was significantly greater than during NIV and SB with a mesh nebuliser (p=0.0001) but not with a jet nebuliser (p=0.384). Delivery efficiency of the mesh nebuliser was greater than the jet nebuliser during MV (p=0.0001), NIV (p=0.0001), and SB (p=0.0001). Aerosol deposition obtained with a mesh nebuliser was greater and differed between MV, NIV, and SB, while deposition was low with a jet nebuliser and similar between the modes of ventilation tested.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Inhaled Medications
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...