Comparing Intra-Arterial, Auscultatory, and Oscillometric Measurement Methods for Arterial Blood Pressure
Keziban Babadağ1, Ayten Zaybak2
1Department of Pediatric Surgery, Ege University, Faculty of Medicine Hospital, İzmir, Turkey.
Aim:
This study aimed to compare the measurement results of arterial blood pressure obtained through intra-arterial, auscultatory, and oscillometric methods.
Method:
This prospective and descriptive study was conducted with 180 patients hospitalized in the intensive care units of cardiovascular surgery and anesthesia. Arterial blood pressures of the patients in the study were measured with 3 methods, and the mean arterial pressure values obtained by each method were analyzed to find out whether they were different or consistent.
Results:
The average systolic blood pressure value using the intra-arterial method was found to be 125.47 ± 21.39 mm Hg, and the average of diastolic blood pressure measurement obtained using the oscillometric method was the highest (73.91 ± 10.62 mm Hg). The highest correlation was seen between the arterial BP measurements of the intra-arterial and auscultatory methods (systolic [0.96] and diastolic [0.90]). According to the British and Irish Hypertension Society protocol, a very good agreement between the diastolic blood pressure values and a good agreement between the systolic blood pressure values were obtained.
Conclusion:
The measurement results obtained through the auscultatory method more consistent with the results obtained through the intra-arterial method compared with those obtained using the oscillometric method.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: