Related Experiment Video
Updated: Oct 28, 2025

Sound Source Localization Testing in Single-sided Deafness Following Bone Conduction Intervention
Published on: December 20, 2024
Bone Conduction Stimulated VEMP Using the B250 Transducer
Karl-Johan Fredén Jansson1, Bo Håkansson1, Sabine Reinfeldt1
1Department of Electrical Engineering, Chalmers University of Technology, Gothenburg, Sweden.
Objective:
Bone conduction (BC) stimulation is rarely used for clinical testing of vestibular evoked myogenic potentials (VEMPs) due to the limitations of conventional stimulation alternatives. The aim of this study is to compare VEMP using the new B250 transducer with the Minishaker and air conduction (AC) stimulation.
Methods:
Thirty normal subjects between 20 and 37 years old and equal gender distribution were recruited, 15 for ocular VEMP and 15 for cervical VEMP. Four stimulation conditions were compared: B250 on the mastoid (FM); Minishaker and B250 on the forehead (FZ); and AC stimulation using an insert earphone.
Results:
It was found that B250 at FM required a statistically significant lower hearing level than with AC stimulation, in average 41 dB and 35 dB lower for ocular VEMP and cervical VEMP, respectively, but gave longer n10 (1.1 ms) and n23 (1.6 ms). No statistical difference was found between B250 at FM and Minishaker at FZ.
Conclusion:
VEMP stimulated with B250 at FM gave similar response as the Minishaker at FZ and for a much lower hearing level than AC stimulation using insert earphones.
More Related Videos
11:06Bilateral Assessment of the Corticospinal Pathways of the Ankle Muscles Using Navigated Transcranial Magnetic Stimulation
Published on: February 19, 2019
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Equilibrium and Balance
The Vestibular System
The Cochlea