Noninvasive continuous arterial pressure monitoring during anesthesia induction in patients undergoing cardiac surgery
Related Experiment Video
Updated: Oct 28, 2025

Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Paul Frank1, Frank Logemann1, Clemens Gras1
1Department of Anesthesiology and Intensive Care Medicine, Hannover Medical School, Hannover, Germany.
Objective:
In this study we compared noninvasive arterial pressure measurement using ClearSight™ vascular-unloading-technique (Edwards Lifesciences Corp, Irvine, CA) with invasive arterial pressure measurement during induction of anesthesia undergoing mayor cardiac surgery.
Design:
Prospective, monocentric.
Setting:
University hospital.
Participants:
54 patients undergoing mayor cardiac surgery.
Interventions:
During induction all patients were simultaneously monitored with invasive (reference method) and noninvasive arterial pressure measurement (test-method) over a mean time period of 27 minutes.
Measurements And Main Results:
We observed slightly lower systolic and mean arterial pressures noninvasive than invasive. For systolic arterial pressure the mean of the differences was -18,05 mmHg (p < 0,05, SD ±16,78 mmHg), the mean arterial pressure MAP -5,47 mmHg (p < 0,05, SD ±11,08 mmHg) and for diastolic pressure -1,09 mmHg (p < 0,05, SD±11,15 mmHg),. The mean of the differences in heartrate was 1,15 (p < 0,05, SD±6,9 mmHg). When considering all measured values of the invasively measured MAP and the ClearSight ™ -MAP at the same timestamp over the recording interval, an almost identical progress can be seen that indicates a sufficient mapping of the hemodynamic changes. The percentage error for mean arterial, systolic and diastolic pressure measured by ClearSight™ amounts to 25,95 %, 26,77 % and 34,16 %, respectively.
Conclusions:
We conclude that ClearSight ™ is a good option for hemodynamic monitoring during induction of anesthesia. Taking into account the limitations, non-invasive arterial blood pressure measurement offers sufficient security to safely initiate anesthesia, especially when MAP is of particular interest. The use of non-invasive arterial blood pressure measurement with ClearSight ™ during induction of anesthesia in patients scheduled for major cardiac surgery is reliable and easy to use.
Related Concept Videos
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Aneurysm IV: Nursing Management
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...