Quantification of image quality of intra-fractional cone-beam computed tomography for arc irradiation with various
Masahide Saito1, Yuki Shibata1, Koji Ueda1
1Department of Radiology, University of Yamanashi, Yamanashi, Japan.
Background:
3-dimensional intra-cone beam computed tomography (intra-CBCT ) could be a potentially powerful tool for use with arc irradiation such as volumetric modulated arc therapy. The aim of the study was to evaluate the image quality of intra-cone beam computed tomography (intra-CBCT ) for arc irradiation with various imaging condition.
Materials And Methods:
Two types of intra-CBCT imaging techniques were evaluated - intra-fractional CBCT with flattening filtered (FF) beam (intra-FF CBCT ) and that with flattening filter free (FFF) beam (intra-FFF CBCT ). For the intra-MV beams, four different field sizes (2 cm × 2 cm, 5 cm × 5 cm, 10 cm × 10 cm, and 20 cm × 20 cm) were used with dose rates of 500 MU/min and 1600 MU/min, for 6 MV FF and 6 MV FFF, respectively. For all image acquisitions, two rotation angles (full-arc and half-arc) were investigated. Thereafter, the linearity, contrast-to-noise ratio (CNR), and uniformity index (UI) of intra-CBCT image were compared with those of conventional CBCT image.
Results:
All acquisition conditions had good linearity of the CT value (R2 > 0.99). For CNR, the change rates from conventional CBCT ranged from 0.6-33.7% for a 2 cm × 2 cm beam, whereas that for a 20 cm × 20 cm beam ranged from 62.7-82.3%. Similarly, the UI increased from 1.5% to 7.0% as the field size increased.
Conclusion:
Quality of intra-CBCT image was affected by the field size and acquisition angle. Image quality of intra-CBCT was worse than that of conventional CBCT, but it was better under a smaller field and wider correction angle and would be acceptable for clinical use.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Radiological Investigation I: X-ray and CT