Related Experiment Video
Updated: Oct 27, 2025

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Magnetic resonance imaging incidents are severely underreported: a finding in a multicentre interview survey
Johan Kihlberg1,2, Boel Hansson3, Annika Hall4
1Department of Radiology in Linköping and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden. johan.kihlberg@liu.se.
Objectives:
The purpose of this study was to develop a procedure to investigate the occurrence, character and causes of magnetic resonance (MR) imaging incidents.
Methods:
A semi-structured questionnaire was developed containing details such as safety zones, examination complexity, staff MR knowledge, staff categories, and implementation of EU directive 2013/35. We focused on formally reported incidents that had occurred during 2014-2019, and unreported incidents during one year. Thirteen clinical MR units were visited, and the managing radiographer was interviewed. Open questions were analysed using conventionally adopted content analysis.
Results:
Thirty-seven written reports for 5 years and an additional 12 oral reports for 1 year were analysed. Only 38% of the incidents were reported formally. Some of these incidents were catastrophic. Negative correlations were observed between the number of annual incidents (per scanner) and staff MR knowledge (Spearman's rho - 0.41, p < 0.05) as well as the number of MR physicists per scanner (- 0.48, p < 0.05). It was notable that only half of the sites had implemented the EU directive. Quotes like 'Burns are to be expected in MR' and not even knowing the name of the incident reporting system suggested an inadequate safety culture. Finally, there was a desire among staff for MR safety education.
Conclusions:
MR-related incidents were greatly underreported, and some incidents could have had catastrophic outcomes. There is a great desire among radiographers to enhance the safety culture, but to achieve this, much more accessible education is required, as well as focused involvement of the management of the operations.
Key Points:
• Only one in three magnetic resonance-related incidents were reported. • Several magnetic resonance incidents could have led to catastrophic consequences. • Much increased knowledge about magnetic resonance safety is needed by radiologists and radiographers.
Insights
Magnetic resonance (MR) imaging incidents are significantly underreported, with only 38% formally documented, some leading to severe outcomes. Enhanced MR safety education and management involvement are crucial for improving safety culture.
Area of Science:
- Medical Imaging
- Radiological Safety
- Healthcare Incident Analysis
Background:
- Magnetic resonance (MR) imaging is a vital diagnostic tool, but associated incidents pose significant risks.
- Understanding the occurrence, character, and causes of MR imaging incidents is essential for patient and staff safety.
Purpose of the Study:
- To develop and implement a procedure for investigating magnetic resonance (MR) imaging incidents.
- To analyze the underreporting of MR incidents and identify contributing factors to inadequate safety cultures.
Main Methods:
- A semi-structured questionnaire was used to collect data on MR safety zones, examination complexity, staff knowledge, and EU directive implementation.
- Analysis included formally reported incidents (2014-2019) and unreported incidents (1 year) from thirteen clinical MR units.
- Open-ended interview questions with managing radiographers were analyzed using content analysis.
Main Results:
- Only 38% of MR incidents were formally reported, with some having catastrophic consequences.
- A negative correlation was found between incident rates and staff MR knowledge, and the number of MR physicists per scanner.
- Inadequate safety culture was indicated by underreporting and a lack of awareness regarding safety protocols and incident reporting systems.
Conclusions:
- MR-related incidents are substantially underreported, with potential for severe outcomes.
- Radiographers express a strong desire for improved MR safety education and a better safety culture.
- Enhanced accessible education and management involvement are necessary to improve MR safety practices.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...