Stopping-power ratios for electron beams used in total skin electron therapy
1Department of Radiation Oncology, Vanderbilt University School of Medicine, Nashville, Tennessee, USA.
Purpose:
The electron beams for total skin electron therapy (TSET) are often degraded by a scatter plate in addition to extended distances. For electron dosimetry, both the AAPM TG-51 and IAEA TRS-398 recommend the use of two formulas developed by Burns et al [Med. Phys. 23, 489-501 (1996)] to estimate the water-to-air stopping-power ratios (SPRs). Both formulas are based on a fit to SPRs calculated for standard electron beams. This study aims to find: (1) if the formulas are applicable to beams used in TSET and (2) the impact of the ICRU report 90 recommendations on the SPRs for these beams.
Methods:
The EGSnrc Monte Carlo code system is used to generate 6 MeV high dose rate total skin electron (HDTSe) beams used in TSET. The simulated beams are used to calculate dose distributions and SPRs as a function of depth in a water phantom. The fitted SPRs using the empirical formulas are compared with MC-calculated SPRs.
Results:
The electron beam quality specifier, the depth in water at which the absorbed dose falls to 50% of its maximum value, R50 , decreases approximately 1 mm for each additional 100-cm extended distance ranging from 2.24 cm at SSD = 100 to 1.72 cm at SSD = 700 cm. For beams passing through a scatter plate, R50 is 1.76 cm (1.14) at SSD = 300 and 1.48 cm (0.85 cm) at SSD = 600 cm with an Acrylic plate thickness of 3 mm (9 mm), respectively. The discrepancy between fitted and MC-calculated SPRs at dref as a function of R50 is <0.8%, and in many cases <0.4%. The difference between fitted and MC-calculated SPRs as a function of depth and R50 is within 1% at depths <0.8R50 for beams with R50 ≥ 1.14 cm. The ICRU-90 recommendations decrease SPRs by 0.3%-0.4% compared to the use of data recommended in ICRU-37.
Conclusion:
The formulas used by the major protocols are accurate enough for clinical beams used in TSET and the error caused using the formulas is <1% to estimate SPRs as a function of depth and R50 for depths <0.8R50 for beams used in TSET with R50 ≥ 1.14 cm. The impact of the ICRU-90 recommendations shows a decrease of SPRs by a fraction of a percent for beams used in TSET.