Related Experiment Video
Updated: Oct 27, 2025

Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
High anatomical accuracy of a novel high-resolution wide-band dielectric imaging system in cryoballoon-based ablation
Laura Rottner1, Milena Nodorp1, Weimann Jessica1
1Department of Cardiology, Universitäres Herz- und Gefäßzentrum Hamburg-Eppendorf, Hamburg, Germany.
Purpose:
Recently, a novel cardiac imaging system based on a wide-band dielectric technology (KODEX-EPD) was introduced to guide catheter ablation. The aim of the study was to evaluate this 3D wide-band dielectric imaging system (WDIS) during cryoballoon (CB)-based atrial fibrillation (AF) ablation focusing on accuracy of pulmonary vein (PV)-anatomy.
Methods:
In consecutive patients with symptomatic AF, CB-based ablation was performed in conjunction with the 3D WDIS. Selective PV-angiographies were performed, and 3D anatomy of the left atrium (LA) and PVs using the 3D WDIS was created. The ostial diameters of the ipsilateral right-sided and left-sided PVs and ostial diameters of the right-/left-sided upper/lower PVs demonstrated by selective angiographies were analyzed and compared to 3D WDIS-based PV visualization.
Results:
In 65 patients (42/65 (65%) male, age 65 ± 9 years, 29/65 (45%) paroxysmal AF) a total of 260 PVs were identified. Median ostial PV-diameters for the ipsilateral left- and right-sided PVs were 38 [34; 43] and 37 [34; 40.3] mm when assessed fluoroscopically and 40 [35.7; 43] and 39 [35.0; 43] mm as demonstrated by 3D WDIS. There was no statistically significant difference between both methods regarding PV-diameter measurements. KODEX-EPD overestimated fluoroscopy measurements by 1.08 mm (95% limits of agreement of -1.93 mm and 4.1 mm).
Conclusion:
The novel wide-band dielectric 3D-imaging system is feasible to create high-resolution images of cardiac structures during CB ablation procedures and accurately visualizes PV-anatomy.