Quantification and Whole Genome Characterization of SARS-CoV-2 RNA in Wastewater and Air Samples
Published on: June 30, 2023
Impact of repeated nasal sampling on detection and quantification of SARS-CoV-2
Joshua M Levy1,2, Jennifer K Frediani2,3, Erika A Tyburski1,4
1The Atlanta Center for Microsystems-Engineered Point-of-Care Technologies, Atlanta, GA, USA.
Abstract:
The impact of repeated sample collection on COVID-19 test performance is unknown. The FDA and CDC currently recommend the primary collection of diagnostic samples to minimize the perceived risk of false-negative findings. We therefore evaluated the association between repeated sample collection and test performance among 325 symptomatic patients undergoing COVID-19 testing in Atlanta, GA. High concordance was found between consecutively collected mid-turbinate samples with both molecular (n = 74, 100% concordance) and antigen-based (n = 147, 97% concordance, kappa = 0.95, CI = 0.88-1.00) diagnostic assays. Repeated sample collection does not decrease COVID-19 test performance, demonstrating that multiple samples can be collected for assay validation and clinical diagnosis.