Reduced serum tricyclic levels due to gel separators
Journal of Clinical Psychopharmacology
|December 1, 1987
Summary
Tricyclic antidepressant levels were lower in serum separator tubes compared to plasma or standard serum tubes. This finding impacts therapeutic drug monitoring for these important medications.
Area of Science:
- Clinical Chemistry
- Pharmacology
- Analytical Toxicology
Background:
- Tricyclic antidepressants (TCAs) are crucial for treating various psychiatric conditions.
- Accurate therapeutic drug monitoring (TDM) of TCAs is essential for optimizing efficacy and minimizing toxicity.
- Different blood collection tube types may influence analyte concentrations.
Purpose of the Study:
- To investigate the impact of different blood collection tube types on serum and plasma tricyclic antidepressant concentrations.
- To compare TCA levels obtained using serum separator tubes (SSTs) versus standard serum tubes and heparinized plasma tubes.
Main Methods:
- Sixty-three blood samples were collected from 48 subjects undergoing TCA therapy.
- Samples were simultaneously drawn into Becton-Dickinson Vacutainer tubes: red stopper (serum, no gel), speckled stopper (serum, gel separator), and green stopper (heparinized plasma).
- Concentrations of imipramine, desipramine, amitriptyline, and nortriptyline were measured.
Main Results:
- TCA concentrations were significantly lower in serum collected with gel separator tubes (SSTs) compared to both plasma and serum from red stopper tubes.
- No significant difference was observed between serum from red stopper tubes and plasma.
- This indicates a potential underestimation of TCA levels when using SSTs.
Conclusions:
- Serum separator tubes may lead to falsely low measurements of tricyclic antidepressants.
- Clinicians and laboratories should be aware of this potential bias in TCA TDM.
- Standard serum tubes or plasma tubes are recommended for accurate TCA level determination.