Measuring peri-implant bone lesions using low-dose cone-beam computed tomography
Laura Christine Leisner1, Akinori Tasaka2, Charlotte Theresa Trebing1
1Department of Prosthodontics, Heidelberg University Hospital, Heidelberg, Germany.
Journal of Prosthodontic Research
|July 26, 2021
Summary
Low-dose cone-beam computed tomography (LD-CBCT) can accurately quantify peri-implant bone lesions, offering a reduced radiation dose. This imaging technique shows minimal differences compared to high-definition CBCT, making it suitable for lesion monitoring.
Area of Science:
- Dental Implantology
- Radiology
- Biomedical Imaging
Background:
- High-definition cone-beam computed tomography (HD-CBCT) provides superior image quality but involves higher radiation doses.
- Low-dose CBCT (LD-CBCT) aims to reduce radiation exposure while maintaining diagnostic accuracy.
Purpose of the Study:
- To evaluate the accuracy of LD-CBCT in quantifying peri-implant bone lesions.
- To compare LD-CBCT measurements with HD-CBCT, considering the influence of surrounding tissues.
Main Methods:
- Titanium implants with peri-implant lesions were imaged using intraoral radiography (IR), HD-CBCT, and LD-CBCT in a human mandible model.
- Measurements of lesion depth and width were performed by four observers and compared to HD-CBCT as the reference standard.
Main Results:
- Measurement deviations from HD-CBCT were minimal for LD-CBCT protocols (depth: 0.22 ± 0.17 mm, width: 0.22 ± 0.13 mm).
- Deviations were slightly larger in LD-CBCT with surrounding tissues simulated (depth: 0.24 ± 0.23 mm, width: 0.25 ± 0.21 mm) and in IR (0.4 mm).
- Excellent interrater agreement for width measurements was observed across HD-CBCT and LD modes.
Conclusions:
- LD-CBCT demonstrates small absolute differences compared to HD-CBCT for peri-implant lesion quantification.
- Despite a decrease in image quality due to surrounding tissues, LD-CBCT shows potential as an adequate imaging modality for monitoring peri-implant lesions at a lower radiation dose.
Keywords:
CBCTDental peri-implantitisDiagnostic accuracyEffect of surrounding tissuesLow-dose protocol