Intact chordae tendineae increase ventricular volume measurement error by the balloon method
1Department of Cardiovascular Dynamics, National Cardiovascular Center Research Institute, Osaka, Japan.
Japanese Heart Journal
|September 1, 1987
Summary
The intraventricular balloon method for measuring left ventricular volume is less accurate with intact chordae tendineae. This method overestimates volume due to space behind the chordae tendineae.
Area of Science:
- Cardiovascular physiology
- Medical imaging techniques
Background:
- Accurate left ventricular volume measurement is crucial for diagnosing and managing cardiac diseases.
- The intraventricular balloon method is a technique used for assessing cardiac volumes.
Purpose of the Study:
- To evaluate the accuracy of the intraventricular balloon method for left ventricular volume measurement.
- To determine the impact of intact versus severed chordae tendineae on volume measurement errors.
Main Methods:
- Utilized formalin-fixed canine left ventricles.
- Assessed volume errors at varying intra-balloon pressures (50-250 mmHg).
- Compared volume errors with all chordae tendineae intact versus severed.
Main Results:
- Volume errors were significantly greater when chordae tendineae were intact compared to when severed.
- The difference in volume errors ranged from 1.9-3.3 ml, irrespective of intra-balloon pressure.
- Visual confirmation indicated the space behind tense chordae tendineae as the primary source of this error.
Conclusions:
- The presence of intact chordae tendineae introduces significant errors in left ventricular volume measurements using the intraventricular balloon method.
- Accounting for the space behind chordae tendineae is essential for improving the accuracy of this measurement technique.