Related Experiment Video
Updated: Oct 26, 2025

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Particle flow rate from the airways as fingerprint diagnostics in mechanical ventilation in the intensive care unit:
Filip Hallgren1, Martin Stenlo2,3,4,5, Anna Niroomand3,4,5,6
1Dept of Cardiothoracic Surgery and Transplantation, Skåne University Hospital, Lund University, Lund, Sweden.
Introduction:
Mechanical ventilation can be monitored by analysing particles in exhaled air as measured by particle flow rate (PFR). This could be a potential method of detecting ventilator-induced lung injury (VILI) before changes in conventional parameters can be detected. The aim of this study was to investigate PFR during different ventilation modes in patients without lung pathology.
Method:
A prospective study was conducted on patients on mechanical ventilation in the cardiothoracic intensive care unit (ICU). A PExA 2.0 device was connected to the expiratory limb on the ventilator for continuous measurement of PFR in 30 patients randomised to either volume-controlled ventilation (VCV) or pressure-controlled ventilation (PCV) for 30 min including a recruitment manoeuvre. PFR measurements were continued as the patients were transitioned to pressure-regulated volume control (PRVC) and then pressure support ventilation (PSV) until extubation.
Results:
PRVC resulted in significantly lower PFR, while those on PSV had the highest PFR. The distribution of particles differed significantly between the different ventilation modes.
Conclusions:
Measuring PFR is safe after cardiac surgery in the ICU and may constitute a novel method of continuously monitoring the small airways in real time. A low PFR during mechanical ventilation may correlate to a gentle ventilation strategy. PFR increases as the patient transitions from controlled mechanical ventilation to autonomous breathing, which most likely occurs as recruitment by the diaphragm opens up more distal airways. Different ventilation modes resulted in unique particle patterns and could be used as a fingerprint for the different ventilation modes.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension