Adverse Events Associated With Electromyogram Endotracheal Tubes in Thyroid and Parathyroid Surgery
Matthew M Pier1, Luke J Pasick2, Daniel A Benito3
1From the Department of Otolaryngology-Head and Neck Surgery, Drexel University College of Medicine, Philadelphia, PA.
Objective:
Thyroid and parathyroid surgeries are frequently performed with the aid of electromyogram endotracheal tubes. However, the most common adverse events have not been reported comprehensively. This study aimed to summarize the device adverse events, patient complications, and subsequent interventions related to electromyogram endotracheal tubes.
Methods:
In this retrospective cross-sectional analysis, the US Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for electromyogram endotracheal tube adverse events between 2010 and 2020. Data were extracted for devices used in thyroid and parathyroid surgery.
Results:
Of 198 included reports, the most common device-related adverse events were loss of response to nerve stimulation (34.8%), cuff perforation (18.2%), and tube lumen obstruction (13.6%). Of 98 patient complications, the most common included ventilation failure (n = 50), airway trauma (n = 22), and recurrent laryngeal nerve injury (n = 5). There was 1 periprocedural death reported. Reported interventions include reintubation performed in 60.1% of all device adverse events, extended hospitalization in 27.3% of airway trauma events, and 5 tracheostomies performed for ventilation failure events. Forty-one events resulted in case delays and 13 events led to aborted cases. The most reported adverse events occurred in 2018 (n = 35) and 2015 (n = 30).
Conclusions:
The most commonly reported adverse events associated with electromyogram endotracheal tubes include loss of response to nerve stimulation, cuff perforation, ventilation failure, and airway trauma. Subsequent interventions include reintubation, tracheostomy, extended hospitalization, and aborted cases. Surgical teams should consider these adverse events when consenting patients and using these devices in thyroid and parathyroid cases.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques