Two-year safety and efficacy of Indigenous Abluminus Sirolimus Eluting Stent. Does it differ amongst diabetics? -
Kamal Sharma1, Sameer Dani2, Devang Desai3
1Department of Cardiology, UNMICRC, BJ Medical College, Ahmedabad (Gujarat), India.
Abstract:
To evaluate the efficacy/safety profile of the Abluminus DES+ over 2-years follow-up in the "real-world" scenario in diabetics as compared to non-diabetics. In prospective, all-comers, open-label registry conducted at 31 sites, patients were analyzed for 1 & 2-year outcomes with the primary endpoint defined as 3P-MACE of CV death, target vessel related myocardial infarction (TV-MI), ischemia-driven target lesion revascularization (TLR)/target vessel revascularization (TVR) apart from Stent thrombosis (ST). Of 2500 patients of PCI with 3286 Abluminus-DES+, 1641 (65.64%) were non-diabetics while859 (34.36%) were diabetics. The 3-P MACE for the cohort at 1 & 2 years were 2.9%, and 3.16%; TLR/TVR - 1.4% at both the intervals for 2493 patients at 2 yrs. follow-up. TV-MI & ST were 0.36% and0.56% at 1st and 2nd year respectively. The 3P-MACE was lower in non-diabetics at 1 & 2 years (2.3%vs 4.2%; 2.4% vs 4.7% respectively). For components of MACE, CV mortality (0.9 vs 1.9% at 1 yr ; 1.0vs 2.1% at 2 years) was significant (P < 0.05) while TLR (1.1 vs 1.9% at 1 yr. & 1.1 vs 2.1% at 2 yrs.) and TV-MI (0.9 vs 1.9% at 1 yr. & 1 vs 2.1% at 2 years) were similar for diabetics and non-diabetics so was ST (P > 0.05). Abluminus-DES+ showed excellent 2-year safety and efficacy with low 3-P MACE which was higher in diabetics driven by higher CV death but similar TLR, TV-MI and ST.
More Related Videos
05:04Implantation of Human-Sized Coronary Stents into Rat Abdominal Aorta Using a Trans-Femoral Access
Published on: November 19, 2020
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023