A new method for anti-negative interference of calcium dobesilate in serum creatinine enzymatic analysis
Hailan Shen1, Kena Chen2, Ju Cao1
1Department of Laboratory Medicine, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.
Background:
Serum creatinine is a widely used biomarker for evaluating renal function. Sarcosine oxidase enzymatic (SOE) analysis is currently the most widely used method for the detection of creatinine. This method was negatively interfered with by calcium dobesilate, causing pseudo-reduced results. The aim of this study was to explore a new method to alleviate the negative interference of this drug on creatinine detection.
Method:
We formulated eight drug concentrations and 12 creatinine concentrations from serum. The SOE method, the new method, and the Jaffe method were used for detection in five systems. Creatinine biases were analyzed under the conditions with or without the interference of calcium dobesilate, at consistent or inconsistent creatinine concentrations. Creatinine concentrations were also analyzed at three medical decision levels (MDLs).
Results:
Calcium dobesilate had negative interference in creatinine SOE analysis. With the increase in calcium dobesilate concentrations, the negative bias increases. The new BG method showed an anti-negative interference effect. In the Roche system, the BG method reduced the negative bias from -71.11% to -16.7%. In the Abbott system, bias was reduced from -45.15% to -2.74%. In the Beckman system, the bias was reduced from -65.36% to -7.58%. In the Siemens system, the bias was reduced from -58.62% to -7.58%. In the Mindray system, the bias was reduced from -36.29% to -6.84%.
Conclusion:
The new method alleviated the negative interference of calcium dobesilate in creatinine SOE detection. The negative bias could be reduced from -60% or -70% to less than -20%.
More Related Videos
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Serum Studies: Renal Function Tests
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...